(1) A licensee providing mobile remote afterloader service for medical use shall perform the following:
    (a) Check survey instruments before medical use at each address of use or on each day of use, whichever is more frequent; and
    (b) Account for all sources before departure from a client’s address of use.
    (2) In addition to the periodic spot-checks required by Fl. Admin. Code R. 64E-5.6421, a licensee authorized to use mobile afterloaders for medical use shall perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of the following:
    (a) Electrical interlocks on treatment area access points;
    (b) Source exposure indicator lights on the remote afterloader unit, on the control console, and in the facility;
    (c) Viewing and intercom systems;
    (d) Applicators, source transfer tubes, and transfer tube-applicator interfaces;
    (e) Radiation monitors used to indicate room exposures;
    (f) Source positioning (accuracy); and
    (g) Radiation monitors used to indicate whether the source has returned to a safe shielded position.
    (3) In addition to the requirements for checks in subsection 64E-5.6423(2), F.A.C., a licensee shall ensure overall proper operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.
    (4) If the results of the checks required in subsection 64E-5.6423(2), F.A.C., indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
    (5) The licensee shall keep a copy of each check for mobile remote afterloader unit required by subsection 64E-5.6423(2), F.A.C., for three years. The records shall include:
    (a) The date of the check;
    (b) The manufacturer’s name, model number, and serial number of the remote afterloader unit;
    (c) Notations accounting for all sources before the licensee departs from a facility;
    (d) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom system, applicators, source transfer tubes, and transfer tube applicator interfaces, and source positioning accuracy; and
    (e) The signature of the individual who performed the check.
Rulemaking Authority 404.022, 404.051, 404.061, 404.071, 404.081, 404.141 FS. Law Implemented 404.022, 404.051(1), (4), (5), (6), (8), (9), (10), (11), 404.061(2), (3), 404.071(1), 404.081, 404.141 FS. History-New 2-11-10.